The Use of a Continuous Glucose Monitoring System (Dexcom G6) in Hospitalized Patients for Acute Care

Overview

The state of a coronavirus disease pandemic (COVID-19) was declared by the World Health Organization on March 11, 2020. Hospitalized patients receiving intravenous (IV) insulin therapy have been identified in the hospital setting as a group that places health care workers at high risk of contamination. Indeed, the many capillary blood glucose measurements required to adjust the IV insulin doses involve direct contact with the patient. It is generally recommended to measure blood sugar every 1 to 2 hours with an IV insulin infusion. Ensuring a working method allowing optimal monitoring of patients' blood sugar levels while limiting the risk of contamination of health professionals is therefore to be preferred. Continuous Glucose Monitoring (CGM) systems are minimally invasive devices that allow you to get a blood glucose measurement every 5 to 15 minutes, and some devices allow you to access your blood sugar profile without having to be in contact with the patients. These devices are approved, validated, and widely used in diabetic ambulatory populations. The principle is based on a measurement of glucose in the interstitial fluid under the skin. DEXCOM G6 is the only device currently approved in Canada that does not require calibration using a capillary blood glucose level or contact with the patient for remote data transmission. A sensor installed on the skin makes it possible to measure blood sugar from the interstitial fluid. A transmitter installed on the sensor sends the data to a receiver which can be remote from the patient (up to 6 meters). Its use in an acute care setting, however, is not approved in Canada. However, Health Canada issued an emergency order on April 25, 2020 allowing the use of Dexcom G6 in hospitals during the COVID-19 pandemic. The CGM offers several potential advantages compared to the capillary glycemia currently used to ensure the follow-up of insulin-treated patients thanks to the real-time measurement of the glycemia and their transmission at a distance. In intensive care the SGC has the potential: – For staff to limit contact with patients infected with COVID-19 (reduction in the use of personal protective equipment and contamination) while ensuring close monitoring of blood glucose levels for the adjustment of insulin therapy. – For patients to reduce the discomfort related to multiple capillary blood sugar levels and possibly better detection of hypoglycaemia This study is the first step in a 2-phase research program, the ultimate goal of which is to use a CGM to replace capillary blood glucose in hospitals to adjust intravenous (IV) insulin therapy. The objective of the first study presented in this protocol is to validate the use of a continuous glucose monitoring system (CGM) for glycemic monitoring in patients hospitalized in intensive care.

Study Type

  • Study Type: Observational
  • Study Design
    • Time Perspective: Prospective
  • Study Primary Completion Date: December 31, 2020

Clinical Trial Outcome Measures

Primary Measures

  • To assess the validity of the Dexcom G6 CGM system measurements in IV insulin therapy patients hospitalized in the intensive care unit of the CHUM compared to the blood glucose measurements made using the Accuchek Inform II glucometer
    • Time Frame: 3-10 days

Secondary Measures

  • To assess the validity of the Dexcom G6 CGM system measurements compared to the different methods of glucose measurement and different blood sources in intensive care
    • Time Frame: 3-10 days
    • Comparing the Dexcom G6 CGM system measurements to the different methods of glucose measurement and different blood sources in intensive care: Portable glucometer, on capillary blood, Portable glucometer, on venous or arterial blood, Central laboratory analyzes on venous or arterial blood
  • Assess the validity of the Dexcom G6 CGM system measurements compared to the glucometer, according to the glucose concentrations measured: <4.0 mmol / L 4.0 – 10.0 mmol / L > 10.0 mmol / L
    • Time Frame: 3-10 days
  • To assess the validity of the Dexcom G6 CGM system measurements compared to the glucometer, according to the rate of change of glucose.
    • Time Frame: 3-10 days
  • Measure the influence of the patient’s clinical characteristics (age, sex, weight, presence of diabetes, septic state and need for organ supports), and biochemical (hematocrit, acidosis, renal function) on the validity of CGM values.
    • Time Frame: 3-10 days
  • Assess the obstacles to the integration of the use of the CGM in the intensive care unit, namely the technical problems encountered and remote data management.
    • Time Frame: 3-10 days

Participating in This Clinical Trial

Inclusion Criteria

  • Adult patients (≥ 18 years old) admitted to intensive care at the CHUM – IV insulin infusion scheduled for a planned duration of at least 72 hours. Exclusion Criteria:

  • Magnetic resonance imaging planned, which would require removing the sensor less than 72 hours after installation; – Patient whose death is expected within the next 72 hours. – Patient included in the LOVIT study due to the theoretical risk of interference with blood glucose measurements with the CMS device

Gender Eligibility: All

Minimum Age: 18 Years

Maximum Age: N/A

Are Healthy Volunteers Accepted: No

Investigator Details

  • Lead Sponsor
    • Centre hospitalier de l’Université de Montréal (CHUM)
  • Provider of Information About this Clinical Study
    • Sponsor
  • Overall Official(s)
    • Rémi Rabasa-Lhoret, MD, PhD, Principal Investigator, Institut de recherches cliniques de Montréal

Clinical trials entries are delivered from the US National Institutes of Health and are not reviewed separately by this site. Please see the identifier information above for retrieving further details from the government database.

At TrialBulletin.com, we keep tabs on over 200,000 clinical trials in the US and abroad, using medical data supplied directly by the US National Institutes of Health. Please see the About and Contact page for details.