Clinical Evaluation of CARESTREAM Dual Energy and Digital Tomosynthesis

Overview

The purpose of this study is to evaluate the imaging performance of Dual Energy and Digital Tomosynthesis (DE-DT) application against the commercially available predicate devices.

Study Type

  • Study Type: Interventional
  • Study Design
    • Allocation: Randomized
    • Intervention Model: Parallel Assignment
    • Primary Purpose: Diagnostic
    • Masking: Double (Participant, Outcomes Assessor)
  • Study Primary Completion Date: April 2016

Detailed Description

The acquisition system will consist of using site equipment including the CARESTREAM DRX-Evolution x-ray system, and the predicate CARESTREAM DRX Plus detector and the investigational DE-DT application. DE Study Design Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue). DT Study Design Fifteen to twenty (15-20) patients will receive a DR standard of care chest exam using the DRX Plus detector, and a DT exam. Each DT patient exam includes a scout image (chest PA) and a DT scan using the investigational DT SW. The DT scan is used by the DT console software to generate tomographic images. 11 phantoms of various anatomy will be imaged with linear tomography (LT) as predicate and DT for investigational.

Interventions

  • Radiation: Radiation
    • Radiation – Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure. Phantoms were imaged using linear tomography and digital tomography.

Arms, Groups and Cohorts

  • Experimental: Predicate & Invest. DE – Human Subjects
    • Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
  • Experimental: Predicate & Invest. DT – Human Subjects
    • Radiation – Fifteen to twenty (15-20) patients will receive a DR standard of care chest exam using the DRX Plus detector, and a DT exam. Each DT patient exam includes a scout image (chest PA) and a DT scan using the investigational DT SW. The DT scan is used by the DT console software to generate tomographic images.
  • Experimental: Predicate & Invest. DT – Phantom Images
    • Radiation – Eleven (11) phantoms of various anatomy will be imaged with linear tomography (LT) as predicate and DT for investigational.

Clinical Trial Outcome Measures

Primary Measures

  • Radlex Scale for Diagnostic Quality Ratings – DE Predicate PA Chest
    • Time Frame: 3 months
    • 1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
  • Radlex Scale for Diagnostic Quality Ratings – DE Investigational Composite
    • Time Frame: 3 months
    • 1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
  • Radlex Scale for Diagnostic Quality Ratings – DE Investigational High Energy
    • Time Frame: 3 months
    • 1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
  • Radlex Scale for Diagnostic Quality Ratings – DE Investigational Low Energy
    • Time Frame: 3 months
    • 1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
  • Radlex Scale for Diagnostic Quality Ratings – DT Reference 1- PA Chest
    • Time Frame: 3 months
    • 1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
  • Radlex Scale for Diagnostic Quality Ratings – DT Reference 2 – PA and LAT Chest
    • Time Frame: 3 months
    • 1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
  • Radlex Scale for Diagnostic Quality Ratings – DT Investigational – Scout & DT Volume
    • Time Frame: 3 month
    • 1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
  • Radlex Scale for Diagnostic Quality Ratings – LT Predicate Phantom Images
    • Time Frame: 3 months
    • 1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
  • Radlex Scale for Diagnostic Quality Ratings – DT Investigational Phantom Images
    • Time Frame: 3 months
    • 1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Participating in This Clinical Trial

Inclusion Criteria

Age 18 years and older. Subject has provided informed consent. Study participant is scheduled to have a chest CT exam. Study participant is able to stand and be still during the exams. Exclusion Criteria:

Human subject specific: Not able or willing to provide Informed Consent, or consent is withdrawn. Study participant is a female of child bearing age.

Gender Eligibility: All

Minimum Age: 18 Years

Maximum Age: N/A

Are Healthy Volunteers Accepted: No

Investigator Details

  • Lead Sponsor
    • Carestream Health, Inc.
  • Provider of Information About this Clinical Study
    • Sponsor
  • Overall Official(s)
    • Narinder Paul, MD, Principal Investigator, Toronto General Hospital